Climate Test Chambers
Climate Test Chamber No. I
Weiss WK 340-70
- Internal dimensions (S x H x D): 540 x 765 x 750 mm
- Distance between editions: ~ 540 mm
- Cable entry: 2 x, left ø 50 mm, right ø 125 mm
Work area temperature:
- Temperature range: -70 °C to +180 °C
- Temperature derivation: ± 0.1 K to ± 0.5 K (temporally), ±0.5 K to ±2.0 K (spatially)
- Cooling speed: 2 K/min in average
- Heat-up speed: 3 K/min in average
Work area climate:
- Temperature range: +10 °C to +95 °C
- Temperature constance: ± 0.1 K to ± 0.3 K (temporally)
- Humidity range: 10 % to 98 % r.H.
- Humidity derivation: ± 1 % to ± 3 % r.H. (temporally)
- Temperature derivation: ± 0.1 K to ± 0.3 K (temporally), ± 0.5 K to ± 1.0 K (spatially)
- Dew point temperature range: + 4 °C to + 94 °C
Climate Test Chamber No. II
Weiss WK 340-70-5
- Internal dimensions (S x H x D): 540 x 750 x 765 mm
- Cable entry: 2x, left ø 50 mm, right ø 125 mm, 1 x front-laterally notching Eexecution 50 x 80 mm
Work area temperature:
- Temperature range: -70 °C to +180 °C, intermittent
- Temperature derivation: ± 0.1 K to ± 0.5 K (temporally), ± 0.5 K to ± 2.0 K (spatially)
- Cooling speed: 6.7 K/min in average
- Heat-up speed: 6.8 K/min in average
Work area climate:
- Temperature range: +10 °C to +95 °C
- Temperature constancy: ± 0.1 K to ± 0.3 K (temporally)
- Humidity range: 10 % to 98 % r.F.
- Humidity derivation: ± 1 % to ± 3 % r.F. (temporally)
- Temperature derivation: ± 0.1 K to ± 0.3 K (temporally), ± 0.5 K to ± 1.0 K (spatially)
- Dew point temperature range: +4 °C to +94 °C
Climate Test Chamber No. III
Vötsch 0060
- Internal dimensions (S x H x D): 800 x 945 x 800 mm
- Cable entry: 1x, 55 mm ø, 1x 125 mm ø
Work area temperature:
- Temperature range: - 5°C bis 90°C, intermittent
- Temperature derivation: ± 0,1 K bis ± 0,5 K (temporally), ± 0,5 K bis ± 2,0 K (spatially)
- Cooling speed: 0,2 K/min in average
- Heat-up speed: 0,5 K/min in average
Work area climate:
- Temperature range: +10 °C to +90 °C
- Temperature constancy: ± 0.1 K to ± 0.3 K (temporally)
- Humidity range: 10 % to 98 % r.F.
- Humidity derivation: ± 1 % to ± 3 % r.F. (temporally)
- Temperature derivation: ± 0.1 K to ± 0.3 K (temporally), ± 0.5 K to ± 1.0 K (spatially)
Climate Test Chamber No. IV
Binder MKF 720
- Internal dimensions (S x H x D): 1000 x 1168 x 600 mm
Work area temperature:
- Temperature range: - 40°C bis 180°C, intermittent
- Temperature derivation: ± 0,1 K bis ± 0,4 K (temporally), ± 0,5 K bis ± 2,5 K (spatially)
- Cooling speed: 3 K/min in average
- Heat-up speed: 3 K/min in average
Work area climate:
- Temperature range: +10 °C to +95 °C
- Humidity range: 10 % to 98 % r.F.
- Humidity derivation: ± 0,5 % to ± 2,5 % r.F. (temporally)
- Dew point temperature range: +5 °C to +94 °C
Climate Test Chamber No. V
Weiss WKL 100/40
- Internal dimensions (S x H x D): 500 x 490 x 400 mm
Work area temperature:
- Temperature range: - 40°C bis 180°C, intermittent
- Temperature derivation: ± 0,3 K bis ± 1,0 K (temporally), ± 0,5 K bis ± 2,0 K (spatially)
- Cooling speed: 3,5 K/min in average
- Heat-up speed: 2,5 K/min in average
Work area climate:
- Temperature range: +10 °C to +95 °C
- Humidity range: 10 % to 98 % r.F.
- Humidity derivation: ± 1 % to ± 3 % r.F. (temporally)
- Dew point temperature range: +5,5 °C to +94 °C


To top